Descripción Técnica

Portátil

Descripción del Anuncio

3 UNITS AVAILABLES

INCLUDED:

- Technical test with certificate of calibration
- 30 days Warranty
- User Manual
- Includes O2 hose
- Includes Bracket
- Includes Patient circuit with valves

PACKAGE: The item is packed and protected

PAYMENT: Bank transfer, Paypal (100% advanced)

RETURN POLICY:
- Return valid for: 30 days after buyer's receipt of the item
- Allowed refund types: Money back, Exchange, Credit Note
- Return shipping & handling will be paid by the buyer
- Additional information: We provide crating and Worldwide

SHIPPING: Please contact us for a quote.

"TAXES NOT INCLUDED"

DESCRIPTION:

• High contrast display
• Oxygen blender for 40 - 100% of O2
• New modes - BIPAP*, ASB, NIV
• Suitable for small children
• Durable for tough, outdoor conditions

TECHNICAL DATA:

Dimensions of basic unit: (W x H x D) 285 x 184 x 175 mm / 11.10 x 7.24 x 6.89 inch (excluding handle)
Weight of basic unit: Approx. 5,4 kg / 11.9 lbs (including internal battery)
Gas supply: From a central gas supply system or from a medical gas cylinder
Supply gas: Medical Oxygen or Medical Air
Supply pressure: 39 - 87 PSI / 270 - 600 kPa at 100 L/min
Gas consumption for internal control: 0.1 to 0.5 L/min

OPERATING DATA:

Ventilation modes IPPV/IPPVassist*, SIMV, SIMV/ASB*, BIPAP*, BIPAP*/ASB*, CPAP, CPAP/ASB*
SPECIAL MODES
• NIV (Non-invasive (mask)-ventilation with leakage compensation)
• Apnea Ventilation (For switching over automatically to volume-controlled mandatory ventilation, if breathing stops)
CPR-mode: Pressure-limited, non-constant-volume ventilation during inspiration time when Pmax is reached
Ventilation frequency: 2 - 60 BPM (SIMV and BIPAP*); 5 - 10 BPM (IPPV*); 12 - 60 BPM (Apnea ventilation)
Tidal volume VT: 50 to 2000 mL, BTPS**
Ventilation time ratio I:E (IPPV*): 1:4 to 3:1
Inspiration time Tinsp. (SIMV, BIPAP*): 0.2 to 10 sec
Inspiratory pressure Pinsp: PEEP + 3 to 55 mbar / cmH2O
O2 concentration: 40 to 100%, infinitely adjustable
PEEP: 0 to 20 mbar / cmH2O
Trigger sensitivity (flow trigger): 3 to 15 L/min
Pressure support ∆ ASB*: 0 to 35 mbar / cmH2O (relative to PEEP), rise time adjustable in 3 steps
Max. inspiratory flow: 100 L/min (supply pressure 350 kPa / 51 PSI), 80 L/min (supply pressure 350 kPa / 51 PSI)
Measured value display: MV, f, VTe, PEEP, Pmean, Ppeak, Pplat, MVspn, fspn, O2***
Display Technology: Electro-luminescence Pixels: 240 * 128, Visible area: 108 * 56 mm / 4.25 x 2.20 inch
Curve display: Pressure curve, flow curve

POWER SUPPLY

Input voltage 19 V DC
AC/DC power pack 100 to 240 V AC / 19 V DC
DC/DC converter 10 to 32 V DC / 19 V DC
Battery type Lithium ion
Operating time (fully charged, Approx. 4 hours (Lithium ion) "typical" ventilation)
Battery charging time Approx 5 hours (Lithium ion)

MONITORING

Supply pressure low: Supply pressure 270 kPa / 39 PSI
Airway pressure (Paw high): Adjustable from 20 to 60 mbar / cmH2O
Airway pressure (Paw low): When pressure difference between Insp. and Exp. 5 mbar / cmH2O or when the set pressure level is not reached
Apnea alarm time TApnea: When respiratory activity is no longer detected, adjustable from 15 to 60s
Leakage (not in NIV): VTexp. is approx. 40% lower than VTinsp.
High frequency: Patient breathes at a high spontaneous rate

OPERATING CONDITIONS

Temperature: - 20 to 50 °C / 14 to 122 °F
Atmospheric pressure: 570 to 1200 hPa / 17 to 35 in. Hg
Rel. humidity: 5 to 95%
Electromagnetic compatibility EMC: In accordance with IEC/EN 60601-1-2:2001 and ISO 10651-3
Airworthiness: In accordance with RTCA DO-160D, sections 7, 8 & 21
Mechanical strength: In accordance with MIL STD 810F, method 514.5
Classification according to MDD 93/42/EEC: Class IIb
UMDNS-Code: 18-098